In the sterilization process in EnbioJet, nutrient medium is pumped by the EnbioJet pump, and then it flows through a teflon pipe section. There, energy coming from microwaves is supplied to the medium. The medium is heated to the temperature of 132 °C, and within several seconds the sterility effect is achieved. The validation performed using the Bacillus Subtillis and Geobacillus stearothermophilus strains confirmed the efficiency of sterilization in 132 °C within 10 seconds.
Sterile nutrient medium flows from the EnbioJet system to the dispensing system. As the process of sterilization in EnbioJet is effected in the flow, simultaneous pouring of the medium is possible. In order to ensure process continuity and stable sterilizer operation, the dispenser should be equipped with a buffer tank. It enables to hold the sterilized nutrient medium in a situation of a momentary dispensing delay.
The equipment has 3 automatic programs:
The EnbioJet sterilizer has its own control panel to set process parameters. The process may be monitored on the screen.
The course of the entire process is stored and available in the form of diagrams. Data is saved on a USB device.